Elster® EnCal 3000 BTU Analyzer

Brand: Honeywell Process Solutions

Honeywell Elster® EnCal 3000 is a gas chromatograph analyzer specially designed for natural gas energy measurements.

This state-of-the-art analyzer uses chromatography components based on the latest Micro-Electro-Mechanical Systems - MEMS.

In natural gas energy measurement, calorific value is a key component that determines the energy content flowing through a pipeline, along with volume, pressure, and temperature measurements.

EnCal 3000 helps gas transmission & distribution companies as well as end-users consuming large amounts of gas accurately measure of single gas components since wrong measurements can directly impact the bottom line.

EnCal 3000 analyzer injects natural gas into a carrier gas stream that further proceeds to a column where natural gas is separated into its constituents.

The separated constituents then pass a calibrated detector in peaks and the area of these peaks is dependent on the constituent’s concentrations.

With collated data from the constituent’s concentrations, the heating value / Wobbe-index / density of the natural gas is calculated
Features
  • C6+ within 3 Minutes
  • C9+ within 5 Minutes
  • Repeatability on BTU < 0.005 %
  • Double Block and Bleed
  • Stream Select for 5 Streams
  • TCP/IP Communication
  • Data Storage in Accordance with API 21.1 Standard
  • IP 66 Outdoor Housing
  • Calculations in Accordance with ISO 6976, GPA 2172 or GOST 22667

Specifications

Analytical Hardware

  • 2 parallel isothermal GC modules with narrow-bore capillary column technology in combination with MEMS based analytical components

Analysis Output

  • Full composition of natural gas up to C6+ or C9+ (option)
  • Heating value, density, Wobbe index

Gas Composition Range

  • N2 : 0 - 22 %
  • CH4 : 55 - 100 %
  • CO₂ : 0 - 20 %
  • C3 : 0 - 10 %
  • C4 : 0 - 10 %
  • Neo-C5 : 0 - 0.25 %
  • C5 : 0 - 0.25 %
  • C6 : 0 - 0.20 %
  • C7 : 0 - 0.10 %
  • C8 : 0 - 0.05 %
  • C9+ : 0 - 0.05 %

Performance

  • Repeatability < 0.005 % for all calculated properties
  • Detection limit for C5: 5 ppm
  • Analysis time: 3 min. for C6+ analysis, 5 min. for C9+ analysis

Ambient Conditions

  • Temperature: -5° to +130° F

Dimensions

  • Base Ø 37 cm x height 37 cm (Ø 14" x height 14")

Weight

  • < 66 lb

Approvals

  • ATEX II2G E Ex d IIB T4, KEMA 05ATEX2191
  • FM Class I, Div. 1 Group B,C and D (Pending)
  • IP 66, Vibration and shock test in acc. with IEC 60068-2-31 and 64
  • EMC according to EN 61000-6-2 and EN 61000-6-4
  • PTB Metrological Certificate Reference No. PTB-3.31-4016861

Power Supply

  • 24 Vdc, 18 W nominal (50 W start-up peak) for non-heated version
  • 24 Vdc, 120 W nominal (170 W start-up peak) for heated version (ambient < 32° F))

Interfaces

  • Ethernet UTP 10 Base-T for ModBus TCP/IP and PC link - Max. distance 330 ft
  • Two RS-232/485 ports for ModBus RTU or ASCII (3-Wire connection for both RS-232 and 485)

Analyzer

  • Complete stand-alone operation, including all calculations and generation of report formats, without need for operator intervention
  • Calculations in acc. with ISO 6976, GPA 2172, GOST 22667 or ASTM D3588

PC Requirements

  • Windows 2000 or Windows XP professional edition (Service Pack 1 or higher)
  • 1000 MHz processor, 512 MB RAM, CD-rom player, free Ethernet port

Data Logging

  • History Log: local storage of last 35 days of all analytical data (analysis, events, alarms, averages, last chromatogram, calibration data) in accordance with API Report 21.1 All data can be made available in XML format.

Sample Gas Inlets

  • Input pressure range 30 - 60 psig, sample gas must be free from particles and liquid, temperature < 130° F .
  • Double block and bleed stream selection for up to 5 streams and 1 calibration gas.
  • Integrated fast loop system with bypass fl ow of 0.7- 1.1 scf/hr. (software selectable)

Helium

  • Quality N5.0, supply pressure 80 ± 7 psig,, consumption ± 1.0 scf/hr
  • Pressure regulator should contain a safety relief set at 95 psig

Calibration Gas

  • Supply pressure 30-60 psig.
  • Consumption ± 1.2 scf/day (at atm. pressure)
  • Composition depending on application

  • Fiscal Energy Measurement
  • Quantitative Measurement of Natural Gas Components
  • Gas Quality Control Measurement